input voltage: | Adaptor ~220V50Hz (85~265VAC,47~63Hz) |
Host DC12V (9~16V) | |
Input power: | No more than 50VA |
Maximum input current: | AC 1A |
DC 2.5A | |
Battery: | 8.4V (2 lithium batteries in series connection) |
Maximum safe pressure of the airway | <10kPa |
System compliance: | Less than 4mL/100Pa。 |
Inspiratory damper | <0.6 KPa @ 60L/min |
Leak | <200 mL/min @ 5Kpa <100 mL/min @ 4Kpa |
Electrical safety: | Comply with the requirements of GB 9706.1-2007 |
Machine noise: | Less than 65dB(A) |
Parameter | range | Resolution/Incremental |
Inspiratory flow rate (Finsp ) | 0 ~ 120 L/min | Continuously adjustable |
Inspiratory oxygen concentration (FiO2) | <45 ~ 100% | 1% |
Respiratory rate(f) | 2 ~ 80bpm | 1 bpm |
Inspiratory timeTI | 0.2 ~ 5.0s | 0.1s |
Triggering pressure(Psens) | -2.0 ~ 2.0Kpa | 0.1Kpa |
The limit pressure of inspiration(Plimit) | 1.0 ~ 5.0Kpa | 0.1Kpa |