YSAV604 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process demonstrate the inner quality: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.
Main body |
High-strength engineering plastic rack, light,beautiful and corrosion resistant |
|
Scope of application |
adult, child |
|
Gas source |
O2: 0.27~0.55MPa N2O: 0.27~0.55MPa |
|
Flow meter |
O2: 0.1~1.0L/min 1.1~10L/min |
|
N20: 0.1~1.0L/min 1.1~10L/min |
||
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
||
Flow rate of rapid oxygen supply |
25~75L/min |
|
Low oxygen pressure alarm |
There will be sound alarm when the oxygen pressure < 0.2MPa |
|
Vaporizer |
it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane,Isoflurane and Sevoflurane,two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
|
Respiratory circuit |
working mode: all-close, semi-close, semi-open APL: 0.5~7 kPa |
|
Respiratory Bellows |
bellows for adults, bellows for children tidal volume range:0~1,500 ml |
|
Ventilator |
||
Display mode |
High-definition 5.7” LCD screen display |
|
Ventilation mode |
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH |
|
Ventilation parameters |
Tidal volume |
50~1500 ml |
Rate |
2~99 bpm |
|
SIMV rate |
2~20 bpm |
|
I:E |
2:1~1:8 |
|
Inspiratory trigger pressure |
-1.0~2.0 kPa |
|
PEEP |
0~2.0 kPa |
|
Pressure range |
1.0~6.0 kPa |
|
SIGH |
1.5 times the inspiratory time |
|
Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate,SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure |
|
Monitoring of oxygen concentration |
(21%~100%) |
|
Security Alarm System |
||
Oxygen Concentration alarm |
upper limit setting range (50%~100%) low limit setting range (15%~50%) |
|
Airway pressure alarm |
upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
|
Per-minute ventilation volume alarm |